- Electrical current applied to motor fibres within nerves causes depolarisation, resulting in contraction of the corresponding muscle(s) and alerting the anaesthetist to the proximity of the nerve
Electrical characteristics
- A nerve stimulator is used to apply DC current (typically 0.2-0.5mA) through an insulated needle, where current density is focused at the needle tip
- The current stimulus is of short duration (0.05 - 1ms) at a frequency of 1-2Hz
- These short stimuli stimulate A⍺-motor neurons and mixed peripheral nerves, but don't stimulate Aδ- or C-fibres and therefore don't trigger pain sensations
- Higher frequencies can cause patient discomfort; lower frequencies risks neural trauma as the needle is advanced between impulses
- The needle outputs a constant, linear current despite changes in resistance of the surrounding tissues
- The needle should be connected to the nerve stimulator via the negative lead (cathode), as this causes an area of depolarisation and requires less current
- If the needle is connected to the positive electrode (anode) then it causes hyper-polarisation near the tip and depolarisation further afield, which requires a greater current to produce the same effect
Nerve localisation
- The higher the current intensity (mA) the larger the 'sphere' of current present at the needle tip and the greater likelihood of causing depolarisation
- The current intensity required to elicit a motor response is inversely proportional to the square of the distance to the nerve (Coulomb's Law)
- As the needle tip is advanced towards the nerve, the current intensity required to cause a motor response reduces exponentially
- Presence of motor response below a certain stimulation threshold (0.5mA) implies high proximity of needle tip to nerve